#*'''Number of exposures:''' The number of exposures to take at the set camera/instrument configuration. '''NB. see also [[#Suitable_Uses|high cadence imaging]] in the Suitable uses section at the top of this page.'''
#*'''Exposure time (in ms):''' Note the exposure time display does not update when the SET button is pushed, it only updates after the exposure is started. This is a [[#Known Issues|known issue]].
#'''Camera Higher Level Instrument Control:'''#*'''Exposure Control:'''<br /> This contains an image display of the last completed exposure, using [https://js9.si.edu/ JS9].
#**'''Take exposure/Start:''' Starts an exposure with the SET exposure configuration
#**'''Abort:''' Aborts the current exposure.
#**'''CCD TemperatureControl:''' Displays the current CCD temperature, and buttons for cooling down and warming up. The camera takes 10 - 15 minutes to cool, and will stabilise at approximately -70 &ordm;C. Cool the camera at the start of observations and warm it up once you are done.
''Note: It is advisable that DARK frames should be taken with the shutter in the "ALWAYS CLOSED" mode. BIAS frames (especially if multiple frames are taken) SHOULD ALWAYS be taken with the shutter in the "ALWAYS CLOSED" mode since this will deactivate the iris-shutter in the closed state and prolong the lifetime, see also [[#Suitable_Uses|high cadence imaging]] in the Suitable uses section at the top of this page.