SPUPNIC

2 bytes added, 14:54, 21 January 2016
/* Focusing the acquisition camera */
===Focusing the acquisition camera===
# * With the guide mirror out of the beam (either set the ''Science'' state in the ''Lamp/Mirror Settings'' pane of the SpUpNIC Control GUI -- if it says Guide Mirror ''Acquisition'', click ''Change'' -- or click on the guide mirror in the instrument schematic to allow the beam through the hole in the guide mirror.# * Turn on the slit illumination, either by clicking on the ''Slit illumination'' icon in the schematic, or by clicking ''Change'' in the ''Instrument Setup'' pane.* On the TCS:# * Start exposures with the acquisition camera and tune the exposure time/LUT sliders/slit illumination to allow good contrast of the scratches on the lower slit jaw.# * Adjust the acquisition camera focus in the bottom right ''Camera Focus'' box until the scratches on the slit jaw appear as sharp as you can get them.
== User's Troubleshooting Guide ==
369