Changes

SPUPNIC

2 bytes removed, 12:33, 7 April 2016
/* Setting up the grating */
===Setting up the grating===
* '''Warning:''' During your start-off you will have been warned told about the grating instability problem. We will be addressing this with a mechanical fix when the instrument is off the telescope in early April. Until such time as we have resolved the problem, we urge users to bracket all target observations with arcs (to be taken immediately before and after the object exposures) in order to minimise the impact of the declination-dependent movement of the grating.
* Note that the wavelength ranges and resolutions listed in the table below are only approximate, the new values are still to be determined from the recent instrument commissioning data!
369
edits