*High-cadence imaging: it is possible to deactivate the shutter and take continues short exposure but users should be aware that the camera is not a frame transfer camera and this mode will result in some readout smearing. There is currently also no GPS triggering of the exposures and that the time recorded in the header (Keyword = DATE-OBS, note: is the start of the exposure) is obtained from the CPU time that is NTP synched. However, the header time has been measured to be accurate within at least 100ms of exact time (see [[Mookodi time-stamp tests|here]]). '''NB. High cadence observing (e.g. multiple continues exposures with exposure times < 60sec) SHOULD NOT be done with the shutter on AUTO mode. It is very important to set the shutter to "ALWAYS OPEN" which will deactivate the shutter in the open state but flush the CCD before the start of each new exposure. Failure to do so will likely cause serious damage to the camera's shutter'''
== Data Reduction Pipelines ===== Spectroscopy ====== Photometry ===
== Operating Mookodi ==
315