1) Log on as ''ccd'' (usual password)
 
2) ''cd'' to working directory
 
3) Enter command: ''new_quartz''
 
4) In the CCD menu, select ''MUS1 ''
 
5) Select ''Local setup files'' → select ''giraffe.gir'' → ''Ready'' → ''Ok''
 
 
'''2.2 Initialising the prisms & filters'''
 
Do this at the start of the run, using options available by clicking on the “Init” button at the top right of the QUARTZ window:
 
1) Select “Init” → “Find reference” and wait for “Finding reference...” window to close
 
2) Select “Init” → “Define red” or “Define blue” depending on the prism you wish to use and enter prism position
 
3) Select “Init” → “Red/blue position go” and wait for “prism movement” and “detector motion” messages to clear
 
4) Check and note down the prism position in the “Prism” status box at the bottom of the QUARTZ window.
 
5) Select “Filter” → “Init” at start of run & before any filter change
 
6) By taking an arc (follow steps 1-2 in Section 4.1), check the resulting orders are correct and spectrograph focus is good (see Section 5).