Difference between revisions of "74" / 1.9m - "Instrument Selector""

From SAAO TOPS Wiki
Jump to: navigation, search
(Instrument Description)
(Instrument Description)
Line 3: Line 3:
 
The "Instrument Selector" is essentially a new fold mirror that has been installed in the acquisition boxes of both SpUpNIC and HiPPO<ref>''22 Feb 2024: the implementation of the fold mirror in HiPPo's acquisition box has not been completed at time of writing</ref> that allows the observer to switch between two instruments: a "primary" and a "secondary" instrument where the primary instrument is either SpUpNIC or HiPPo and the secondary instrument is always a new SHOC system. In other words, a SHOC is from now on permanently available on the 74-inch.  
 
The "Instrument Selector" is essentially a new fold mirror that has been installed in the acquisition boxes of both SpUpNIC and HiPPO<ref>''22 Feb 2024: the implementation of the fold mirror in HiPPo's acquisition box has not been completed at time of writing</ref> that allows the observer to switch between two instruments: a "primary" and a "secondary" instrument where the primary instrument is either SpUpNIC or HiPPo and the secondary instrument is always a new SHOC system. In other words, a SHOC is from now on permanently available on the 74-inch.  
  
Below is the image showing the new SHOC system mounted to the side of SpUpNIC's acquisition box. The new SHOC system replaces the old guider box that use to migrate between SpUpNIC and HiPPo. The new SHOC system is incorporated into a newly designed off-axis "pick-off" style guider system that can accommodate a filter wheel unit and SHOC camera behind the guider. The new SHOC system also has a focal reducer built in providing roughy double the FoV width compares to the the old system.
+
Below is the image showing the new SHOC system mounted to the side of SpUpNIC's acquisition box. The new SHOC system replaces the old guider box that use to migrate between SpUpNIC and HiPPo. The new SHOC system is incorporated into a newly designed off-axis "pick-off" style guider system that can accommodate a filter wheel unit and SHOC camera behind the guider. The new SHOC system also has a focal reducer built in providing roughy double the FoV-width compared to the the old system.
  
 
To select between the primary instrument and SHOC, a user can actuate the fold mirror in/out of the beam to direct the science beam towards the desired instrument (see Figure 2 below that indicates the button on TCS that actuates the fold mirror). When the fold mirror is IN beam and the guider's XY stages are "centred" then the guider camera also functions as a "target acquisition" camera for SpUpNIC/HiPPo, similar to the old system. Also as in the old system, the guider can patrol the area of the large annulas mirror to acquire an appropriate guide star.
 
To select between the primary instrument and SHOC, a user can actuate the fold mirror in/out of the beam to direct the science beam towards the desired instrument (see Figure 2 below that indicates the button on TCS that actuates the fold mirror). When the fold mirror is IN beam and the guider's XY stages are "centred" then the guider camera also functions as a "target acquisition" camera for SpUpNIC/HiPPo, similar to the old system. Also as in the old system, the guider can patrol the area of the large annulas mirror to acquire an appropriate guide star.

Revision as of 00:55, 23 February 2024

Instrument Description

The "Instrument Selector" is essentially a new fold mirror that has been installed in the acquisition boxes of both SpUpNIC and HiPPO[1] that allows the observer to switch between two instruments: a "primary" and a "secondary" instrument where the primary instrument is either SpUpNIC or HiPPo and the secondary instrument is always a new SHOC system. In other words, a SHOC is from now on permanently available on the 74-inch.

Below is the image showing the new SHOC system mounted to the side of SpUpNIC's acquisition box. The new SHOC system replaces the old guider box that use to migrate between SpUpNIC and HiPPo. The new SHOC system is incorporated into a newly designed off-axis "pick-off" style guider system that can accommodate a filter wheel unit and SHOC camera behind the guider. The new SHOC system also has a focal reducer built in providing roughy double the FoV-width compared to the the old system.

To select between the primary instrument and SHOC, a user can actuate the fold mirror in/out of the beam to direct the science beam towards the desired instrument (see Figure 2 below that indicates the button on TCS that actuates the fold mirror). When the fold mirror is IN beam and the guider's XY stages are "centred" then the guider camera also functions as a "target acquisition" camera for SpUpNIC/HiPPo, similar to the old system. Also as in the old system, the guider can patrol the area of the large annulas mirror to acquire an appropriate guide star.

Figure2: Button (and indicator) that actuates the fold mirror. It will remain green in both states (IN or OUT)

Important Notes

NB. (1) Because the fold mirror and annulas guide mirror are displaced relative to one another in the optical z-axis the guide/acquisition camera's focus has to be adjusted substantially when switching between guiding and target-acquiring mode. This cheat-sheet provides guidelines for focus settings and summarises instructions how to set up target-acquisition and guiding for the primary and secondary instrument: cheat-sheet.

NB. (2) The annulas mirror, which used to be actuated to switch between acquistion and guiding mode, should NOT be actuated anymore for any reason. There is no danger in doing so but the new routines rely on the annuls mirror staying in the "SCIENCE" position (see Figure 3 below). There are plans to maybe disable the actuating motor eventually.

Figure3: The annulas "Guide Mirror" should always remain in the "SCIENCE" position as indicated and should NOT be actuated anymore for any reason.
  1. 22 Feb 2024: the implementation of the fold mirror in HiPPo's acquisition box has not been completed at time of writing