Changes

SHOC

65 bytes added, 12:49, 14 June 2016
/* LED timing tests */
''Currently, this functionality is only available on the 74". Eventually, it should also be on the 40" and the new 1m.''
*There is an LED mounted in the telescope optical path that can be used to independently verify SHOC timing. Log into the following website:
http://10.2.55.5:4000/timepulse
Ask the Electronics standby for username and password. This website allows selection of LED pulses of different cadence (1, 2, 5, 10, 100 Hz and 100kHz), see the image below. '''Note that the pulse must be switched off when you are done, otherwise it will continue blasting photons into the telescope!'''
 [[image:74inLEDPulseWebsite.jpg|500px]] *Instructions on what data to take and how to analyze will be posted soon.
== In-depth Information ==
214
edits