SHOC User Manual
NOTE: THIS PAGE REFERS TO THE LINUX- and WEB-BASED SOFTWARE (AS OF MARCH, 2015). For the original, manufacturer-based software and instrument commissioning details, please refer to the manual at http://shoc.saao.ac.za/Documents/ShocnHelpful.pdf.
- If you have found this webpage already loaded on the browser, please refresh to ensure that it is current.
Introduction
The SHOC systems consist of a filterwheel unit (with one or two filterwheels), a GPS unit for providing accurate timing triggers, an Andor iXon camera, and a control PC. Currently there are 2 full SHOC systems (shocnawe & shocndisbelief), and 1 development system running a different camera (shocnhorror). As of March 2015, the control software is web-based. Users point their browser to the instrument location, and separate tabs appear to allow control of the camera, GPS, and filter wheel. SHOC can be mounted on the 1.9- and 1-m telescopes. We have tried to make the usage telescope specific, but it is very useful to know which SHOC instrument is mounted on the telescope you are using. Details on instrument operation are below.
The SHOC website may well have more extensive detail about this instrument.
Dr. Marissa Kotze has written some pipeline software which may help with reduction of SHOC data.
IMPORTANT NOTES ABOUT THE NEW SOFTWARE
- THE NEW SOFTWARE HAS DIFFERENT FILE NAMES AND HEADER KEYWORDS THAN PREVIOUS DATA. If you use this information in your analyses, you will have to update it. The filename format has been standardized to <Instrument>_YYYYMMDD.####.fits, where the instrument is SHA, SHD, or SHH to designate which SHOC box is being used. Please check the header for the keywords (the "new" keywords are directly from the Andor iXon camera under Linux).
- THE FILE SIZE IS NO LONGER LIMITED TO 2 GB. This means that if you take large data cubes, the file size has no limit (this is because we are no longer on a Window-based, 32-bit, operating system). If you end up with a very large file, data analysis can be difficult. We are working on an efficient method to automatically split files into smaller sizes, but for now users need to be aware that requesting a large number of images in one file could return a very large file size (and if you mistakenly delete your file, you have lost alot of work!). It is recommended that multiple files be manually taken. If you have any questions or problems with the data, feel free to contact carel@ saao.ac.za or amanda@ saao.ac.za.
- If you use the GPS, please set the pulse width to 10 microsec. It is currently defaulting to 1 microsec, and this will be fixed in the next version of the software.
Change Log
v1.1
Quick Start
- (1) Connect to the web browser. Go to http://shoc40in.suth.saao.ac.za:5000 or http://shoc74in.suth.saao.ac.za:5000 depending on which telescope you are using.
- (2) Initialize the filter wheels and select your filter. You must initialize both filter wheels whenever the webservice is started.
- (3) Click on camera tab: turn on the camera and confirm that it is cooling.
- (4) If using the GPS to trigger data frames, click on GPS tab: check that the GPS has valid time and can be set.
- (5) Click on camera tab: set parameters on the camera and start previewing and taking data.
- (6) Transfer data from the instrument to your local machine or the SAN.
- (7) Turn OFF the camera before you leave!!
Filter wheel
- (1) Point your browser at shoc40in.suth.saao.ac.za:5000 or shoc74in.suth.saao.ac.za:5000
NOTE: ON DNS ALIASES: These telescope web addresses are aliases to the SHOC instruments. The real addresses for the instruments are shocnawe.suth.saao.ac.za:5000 or shocndisbelief.suth.saao.ac.za:5000. In order to avoid confusion, or the requirement that the user know which instrument is mounted on their telescope, we have created the aliases.
- (2)Select the filterwheel tab
- Initialize the filterwheel(s) (3), then select the filter required (4), click "Move" (5). The "Centered" indicator should be green when the wheel is finished moving (6).
GPS
- If using external triggering, select the GPS tab (1)
- Ensure timing is valid (2).
- Set the Programmed Output Pulse (POP) parameters (mode (3), pulse width (4), start date (5), start time (6), trigger interval (7)) and click apply (8). The status should change to pending (9).
- Note that the current default pulse width is 1 microsec and it should be 10 microseconds. Please change it!
Camera
- Select the camera tab (1).
- Turn the camera on (2)
- Set the desired Readout Rate (1), Preamp (2) and Electron Multiplying (3) parameters (NB: EM MODE MAY ONLY BE SET IF YOU HAVE WRITTEN PERMISSION TO DO SO).
- Set the sub image (4) and binning (5) parameters.
- Set the trigger type (6), exposure (7) and kinetic series (8) parameters.
- Start the acquisition (9).
Acquiring data
- Stop to abort (1), or wait until completion. The image display area (based on DS9) (2) shows snapshots of the images throughout the acquisition.
- When the series is complete, the FITS data cube is copied to /data/<telescope>/<INSTR>/<YYYY>/<MMDD>/<filename>.fits, where:
- <telescope> is 40in or 74in,
- <INSTR> could be SHD for shocndisbelief, SHA for shocnawe or SHH for shocnhorror,
- <YYYY> is the year,
- <MMDD> is the month and day, (Note that YYYY/MMDD is that of the start of the night and is increased at NOON the following day.)
- <filename> is the instrument abbreviation (SHA or SHD) followed by the date in YYYYMMDD format and the number of the acquisition.
- An example is /data/74in/SHA/2015/0325/SHA_20150325.0001.fits
- Make sure to turn off the camera when you are done for the night.
Data Transfer
- To transfer data directly to your computer from the SHOC instrument, use the command line. Copy the data using the following command, depending on the telescope:
rsync -avzP shoc74in@shoc74in.suth.saao.ac.za:/data/74in/SH?/<YYYY>/<MMDD>/*.fits /LocalFilePath rsync -avzP shoc40in@shoc40in.suth.saao.ac.za:/data/40in/SH?/<YYYY>/<MMDD>/*.fits /LocalFilePath
NOTE: Data are moved to the SAN on the plateau every morning at 9:30am. However, if you're wanting to get your data to the SAN (and from there to the servers in Cape Town sooner than 9:30am) you MUST have run the shocdatacopy.sh script in order to get the data to the SAN. You can then obtain the data from astro.suth.saao.ac.za (or astro.cape.saao.ac.za if you're back in Cape Town). See Data Transfer information under the advanced guide below for details on how to transfer to the SAN.
Advanced users guide
Web interface
- The primary means of controlling SHOC is via a web interface. The Firefox or Chrome browsers are preferred.
- Point your browser at http://shoc74in.suth.saao.ac.za:5000 or http://shoc40in.suth.saao.ac.za:5000
- The interface should appear, showing three tabs: Filterwheel, GPS and Camera.
Filter wheel
- Select the filterwheel tab. There may be one or two wheels physically present; the interfaces to these are displayed side-by-side. The interface values for a wheel will be empty if the wheel is absent.
- Initialize the wheel by clicking the "Initialize" button (usually only at the start of the night, but may also be done if you suspect that the controller has lost its way). The wheel will move around to the reference position; this may take several seconds. While still moving, the "Moving" indicator above the controls will be red. Once there, the "Initialized" and "At Ref" indicators should turn green. The value in the "Current Position" field should be 1 after initialization. YOU MUST INITIALIZE BOTH FILTER WHEELS AT THE START OF THE NIGHT, AND MOVE ONE TO AN "EMPTY" POSITION TO ENSURE THAT THE CORRECT FILTER IS IN THE BEAM.
- Select the filter required, by using the up and down arrows on the right of the "Required Position" field. The filter name will indicate the filter in the selected position. Click the "Move" button. The indicators above the wheel controls should change; once the wheel has reached the required position, the "Moving" indicator should turn grey, while the "Centred" indicator should be green. The "Current Position" field will indicate the slot number and name of the filter now in the light path.
GPS (optional: to be used for accurately-timed images)
- The GPS tab allows the user to view the time, GPS state, and to change the Programmed Output Pulse (POP) settings.
- Check that the time is valid, and that there are no alarm conditions indicated.
- Make sure the pulse width is set to 10 microsec. It might function with shorter pulses, but there is a chance they could be missed. We typically use 10 microsec as the default.
- To set the Programmed Output Pulse, choose whether you want a single or repeating pulse, then choose a pulse width, the start date and time (in local time, 24 hour format), and for repeating pulses, the repeat interval. Then push the "Apply" button. Note: the POP may only be set to occur in the future; if an error occurs, check that you haven't chosen a time in the past.
- To stop the pulses from being sent, click the "Stop" button.
Camera
- The camera tab allows control and setting of the SHOC imaging camera.
- Before using the camera, you need to switch it on: click the "Turn Camera On" button.
- Once the camera has been initialized, the camera setting and view interface is displayed.
- On the left is the image display area; more on that later.
- On the right is the settings area, with three tabs for camera control, analysis and advanced settings.
- When switched on, the camera starts cooling to its target temperature. The temperature display button is orange while the camera is not yet at the target temperature, and green once it is. If using a non-standard temperature, this may be set using the "Advanced" tab.
- The camera may be used in internal, external or external start modes, selectable using the "Triggering" drop-down.
- While in internal or external start trigger modes, the exposure time may be set, and the kinetic cycle time is automatically calculated and displayed. The latter is the period between the start of successive frames in a kinetic series. Arbitrarily short exposre and kinetic cycle times are not possible, due to the time required to read out the image. If you choose an exposure time less than the minimum allowed, the minimum exposure and kinetic cycle times are calculated and automatically displayed.
- In external trigger mode, the kinetic cycle time is determined by the frequency of the GPS POP signals. However, this frequency should not be higher than than the minimum readout time. The minimum allowed kinetic cycle time is calculated and displayed; this may be useful when deciding on the GPS POP frequency to use.
- For all trigger modes, the number of frames in the data cube is determined by the Kinetic Series value.
- The next row down allows the sub-image area and binning to be chosen. For both of these, standard or custom values may be used. Smaller sub-images and higher binning allow faster image readouts, at the expense of image size and resolution. The minimum values discussed above are automatically re-calculated and displayed when the sub-image or binning is changed. NOTE: BINNING DIMENSIONS SHOULD BE EQUALLY DIVISIBLE INTO THE AREA OF THE SUB-IMAGE WHEN USING CUSTOM VALUES. An error message will appear, and will need to be cleared, when the values are not evenly divisible.
- You can visually draw a custom sub-image in the display area by clicking the white square to the right of the sub-image options. This will display a red region which can be manipulated to select the desired sub-image. Click the square again to hide the region in the display area. NOTE: YOU WILL NEED TO CAPTURE A FRAME BEFORE YOU CAN SELECT A SUB-FRAME USING THIS METHOD.
- Below this, the readout speed and preamp gain may be set. Higher readout speeds allow for shorter exposure times, but at the expense of noisier data.
- The last setting is the Electron Multiplying (EM) gain. NOTE: DO NOT USE ELECTRON MULTIPLYING MODE UNLESS YOU HAVE SPECIFICALLY OBTAINED PERMISSION FROM DR AMANDA GULBIS TO DO SO, AND YOU UNDERSTAND HOW AND WHEN TO USE IT. PERMANENT DAMAGE TO THE SENSOR CAN RESULT FROM INCORRECT USAGE. If you are using EM mode, choose the EM output amplifier, and then select a gain. BEGIN WITH LOW GAIN VALUES, BECAUSE IF THERE IS TOO MUCH SIGNAL YOU CAN DAMAGE THE CAMERA.
- To get a quick view of the images that will result, press the "Preview" button. Images will be displayed in the left-hand display panel. Press "Stop" when you're done.
- To begin collecting data, press the "Start" button. The kinetic series will start, either immediately or at the designated time if using the GPS, and the data will be spooled to disk. At the same time, temporary images will be displayed in the image display area.
- Once the acquisition is finished (or earlier, if you push the "Stop" button), the header information in the FITS cube will be updated to include various GPS, filter and environmental parameters, an the cube written to disk. The filename for the current or past kinetic series is displayed above the "Start" button. The filename consists of the instrument name, the date and an index. The index is updated every acquisition. and the new filename is calculated and displayed immediately after the "Start" button is pressed.
- The index appended to file names begins at 1, and increments by 1 after each acquisition. It may be edited by clicking on the "Advanced" tab, and changing its value. Note that if you change it to a value already used, it will cause files with that value to be overwritten.
- For convenience, the following additional keywords can be inserted into your data files from the "Advanced" tab: Observer, Observation Type, Object, Right Ascension, Declination, Epoch and Equinox.
- You also have the ability to flip the image orientation along the X and Y axes. These options are provided in the "Advanced" tab too.
* Once you're finished with the camera, remember to turn it off! It should not be left with the cooler on once the night's observations are finished.
Real-time image display
- As an acquisition proceeds, snapshot images are displayed in the display area of the camera tab, basically as fast as the browser can render them.
- The interface is based on JS9, a Javascript version of the DS9 FITS processing and display engine (so users familiar with DS9 should be able to adapt to the JS9 interface relatively easily).
- Menu items above the image allow users to change the image zoom level, scaling and colour. It is also possible to select regions of the graph for further analysis.
Data Transfer
There are two ways to transfer data: (1) directly to your computer or (2) onto the SAAO storage area network (SAN). The data are automatically copied to the server every morning (at 9:30am).
- (1) To transfer data directly to your computer, use the command line.
- Open a shell on your local PC (Putty if you're using Windows or a bash/tcsh/sh if on Mac or Linux)
- Copy the data using the following command, depending on the telescope:
rsync -avzP shoc74in@shoc74in.suth.saao.ac.za:/data/74in/SH?/<YYYY>/<MMDD>/*.fits /LocalFilePath rsync -avzP shoc40in@shoc40in.suth.saao.ac.za:/data/40in/SH?/<YYYY>/<MMDD>/*.fits /LocalFilePath ---- where SH? is a wildcard that will point to the correct instrument mounted on the telescope for that night and "/LocalFilePath" is the location on your local machine ---- the password is that for the shoc74in or shoc40in accounts
- (2) To move data to the SAN on the Sutherland plateau (from which you can access it by connecting to the SAAO server), there is a data copy script on the SHOC computer.
- From a command line, connect to the instrument computer. As appropriate for your telescope, use either of the following commands:
ssh shoc74in@shoc74in.suth.saao.ac.za ssh shoc40in@shoc74in.suth.saao.ac.za ---- When prompted for the password, use the password for the 74in or 40in user.
- Run the shocdatacopy.sh script. Options can be:
shocdatacopy.sh -d yesterday # to transfer yesterday's data shocdatacopy.sh -d YYYYMMDD # to transfer data from some specified date ---- Without parameters, it will try to transfer today's data.
- To collect your data from the SAN, you'll have to secure copy it from astro.suth.saao.ac.za. The usernames for this are ccd40 and ccd74 depending on your telescope. Ask for the password. Data are in the path /data/telescopedata/<TELESCOPE>/SH?/<YYYY>/<MMDD>
rsync -avzP ccd74@astro.suth.saao.ac.za:/data/telescopedata/74in/SH?/<YYYY>/<MMDD>/*.fits /LocalFilePath rsync -avzP ccd40@astro.suth.saao.ac.za:/data/telescopedata/40in/SH?/<YYYY>/<MMDD>/*.fits /LocalFilePath ---- where SH? is a wildcard that will point to the correct instrument mounted on the telescope for that night; YYYY is the year; MMDD is the month and day; and "/LocalFilePath" is the location on your local machine
- Useful:
- On the SHOC computers, the 'tree' command will assist you in trying to look at what folders and files are under that directory.
- For information on the script, a logfile is produced that will be in /var/log/<TELESCOPE>DataCopyLog<MMDD>-<NUM> where <TELESCOPE> is 40in or 74in <MMDD> MonthDay <NUM> the Nth time this script has been run
Rescuing Data
Under normal operation, the system automatically updates the FITS cube header with supplementary information (filter, external trigger parameters, telescope pointing information etc.) and the moves the cube to the permanent (for the night) storage area. From time to time, it may be necessary to do this manually, using the rescue_data.py tool. In particular, if the system has frozen at the end of an acquisition, but before fixing the header and moving the data file, it will be necessary to perform this procedure. This procedure does NOT need to be done when stopping an acquisition, only if the interface hangs and you cannot do anything.
- The procedure is to open a terminal and find your data in the spool directory on the instrument computer. It's stored in a directory named /data/spool/<CCYYMMDD><index>.<HHMMSS> (where index is the index, CCYYMMDD is the date and HHMMSS is the time the acquisition started). For example, /data/spool/201503250002.141026/ . The procedure is done by using ssh to connect to the SHOC host computer:
ssh shoc74in@shoc74in.suth.saao.ac.za
or
ssh shoc40in@shoc74in.suth.saao.ac.za
When prompted for the password, use the password for the 74in or 40in user. Then go to the directory containing the spooled files:
cd /data/spool ls -l cd <CCYYMMDD><index>.<HHMMSS> ls -l
- There should be files named data.fits and info.txt
- The info.txt file contains the supplementary information.
- Run the rescue utility: type
sudo rescue_data.py
and then answer the questions as prompted. You'll be prompted for a password, use the password for the user you logged in as.
- The telescope is the name of the telescope being used; either 40in or 74in.
- The instrument will be one of shocnawe (SHA) or shocndisbelief (SHD). These are used to create the directory and filename in the /data directory. The directory will be of the form /data/<telescope>/<instrument>/<YYYY>/<MMDD>, and the file of the form <instrument>_<YYYYMMDD>.<index>.fits. From this, the instrument name can be determined. At the moment, SHA is the SHOC instrument mounted on the 74inch.
- You will be prompted for an index to use. NB: make sure there is no existing file with the same index that will be overwritten The script will scan the target directory and list the existing files. Choose and index which isn't used(one option is to use an index like 3a , if index 3 exists). In other words, if files such as SHA_20150403.0004.fits and SHA_20150403.0005.fits, choosing either index 4 or 5 will result in that file being overwritten. If you choose as index 4a or 5a, the files will not be overwritten.
FITS Header Keywords
The FITS information is mostly drawn from the Andor camera. We have added additional keywords added to contain information on target (which can be entered by the user under the "advanced" tab in the web interface), the GPS settings, and from the telescope (when available -- currently only for the 74 inch). We have retained the raw state of Andor's keywords in order keep things as simple and straightforward as possible.
Keyword and comments (example) | Notes |
---|---|
SIMPLE = T / file does conform to FITS standard | |
BITPIX = 16 / number of bits per data pixel | |
NAXIS = 3 / number of data axes | |
NAXIS1 = 128 / length of data axis 1 | |
NAXIS2 = 128 / length of data axis 2 | |
NAXIS3 = 10 / length of data axis 3 | |
EXTEND = T / FITS dataset may contain extensions | |
COMMENT FITS (Flexible Image Transport System) format is defined in 'Astronomy | |
COMMENT and Astrophysics', volume 376, page 359; bibcode: 2001A&A...376..359H | |
BZERO = 32768 / offset data range to that of unsigned short | |
BSCALE = 1 / default scaling factor | |
HEAD = 'DU8201_UVB' / Head model | |
ACQMODE = 'Frame Transfer' / Acquisition mode | |
ACT = 1.11592 / Integration cycle time | |
KCT = 1.11592 / Kinetic cycle time | |
NUMACC = 1 / Number of integrations | |
NUMKIN = 10 / Series length | |
READMODE= 'Image ' / Readout mode | |
IMGRECT = '1, 1024, 1024, 1' / Image format | |
HBIN = 1 / Horizontal binning | |
VBIN = 1 / Vertical binning | |
SUBRECT = '897, 1024, 128, 1' / Subimage format | |
DATATYPE= 'Counts ' / Data type | |
XTYPE = 'Pixel number' / Calibration type | |
XUNIT = 0 / Calibration units | |
RAYWAVE = 0. / Rayleigh Wavelength | This seems to be irrelevant, since we are not taking spectra. |
CALBWVNM= 0 / Wave calibration | This seems to be irrelevant, since we are not taking spectra. |
TRIGGER = 'Internal' / Trigger mode | |
CALIB = '0,1,0,0 ' / Calibration | |
DLLVER = '2.97.30005.0' / Software Version | |
EXPOSURE= 1.10916 / Total Exposure Time | |
TEMP = -999. / Temperature | If camera temp is not stabilized, this will be -999
and the temperature will be under keyword "UNSTTEMP". |
READTIME= 1.0E-06 / Pixel readout time | This corresponds to the 1, 3, 5, or 10 MHz readout amplifier. |
OPERATN = 4 / Type of system | |
GAIN = 0 / Gain | |
EMREALGN= 0 / EM Real Gain | |
VCLKAMP = 0 / Vertical Clock Amplitude | |
VSHIFT = 6.5E-06 / Vertical Shift Speed | |
OUTPTAMP= 'Conventional' / Output Amplifier | |
PREAMP = 1. / Pre Amplifier Gain | This corresponds to the 1, ~2.5x, or ~5x gain setting.
See keyword SNTVTY or the camera spec sheets at http://shoc.saao.ac.za/Documents/iXon.X5982.SpecSheet.pdf and http://shoc.saao.ac.za/Documents/iXon.X6448.SpecSheet.pdf. |
SERNO = 6448 / Serial Number | The serial number of the camera. |
UNSTTEMP= -35.764 / Unstabilized Temperature | If camera temp is stabilized, this will be -999
and the temperature will be under keyword "TEMP". |
BLCLAMP = F / Baseline Clamp | |
PRECAN = 0 / Prescans | |
FLIPX = 1 / Horizontally Flipped | |
FLIPY = 0 / Vertically Flipped | |
CNTCVTMD= 0 / Count Convert Mode | |
CNTCVT = 1 / Count Convert | |
DTNWLGTH= 500. / Detection Wavelength | A misleading keyword: the camera has no information about filters! |
SNTVTY = 3.82 / Sensitivity | The e/ADU conversion. |
SPSNFLTR= -5551 / Spurious Noise Filter Mode | A keyword that we are still trying to decipher. There are multiple camera
settings, and they result to values ofeither 0 or -5551. Setting 0 returns normal-looking data, but returns this -5551 value. |
THRSHLD = 0. / Threshold | |
PCNTENLD= 116 / Photon Counting Enabled | A keyword that we are still trying to decipher.
When this camera setting is "off" the value of 116 is listed in the header. When the setting is "on" the value is 0. |
NSETHSLD= 0 / Number of Photon Counting Thresholds | |
PTNTHLD1= 0. / Photon Counting Threshold 1 | |
PTNTHLD2= 0. / Photon Counting Threshold 2 | |
PTNTHLD3= 0. / Photon Counting Threshold 3 | |
PTNTHLD4= 0. / Photon Counting Threshold 4 | |
AVGFTRMD= 0 / Averaging Filter Mode | |
AVGFCTR = 1 / Averaging factor | |
FRMCNT = 1 / Frame Count | |
USERTXT1= ' ' / User text | |
USERTXT2= ' ' / User text | |
USERTXT3= ' ' / User text | |
USERTXT4= ' ' / User text | |
DATE = '2015-04-17T14:34:28' / file creation date (YYYY-MM-DDThh:mm:ss) | |
FRAME = '2015-04-17T14:34:28.000' / Start of Frame Exposure | This comment is incorrect. It is the time at the end of the first frame
for internal mode, and the time "start" is pressed for external triggering mode. |
ESHTMODE= 0 / Electronic Shuttering Mode | This is the last keyword of those coming straight from the Andor Ixon camera. |
AIRMASS = 'NA ' / The airmass (sec(z)) | Starting with this keyword, everything here and below has been
added in the SAAO software. Airmass comes from the TCS. |
DATE-OBS= '2015-04-17T14:34:26.446239' / The time the user pushed start (UTC) | Note that during external triggering mode, this does not correspond to
any specific timing for the images. The camera started waiting at this time for a trigger from the GPS. |
DOMEPOS = 'NA ' / The dome position | Read from the TCS. |
FILTERA = ' ' / The active filter in wheel A | |
FILTERB = ' ' / The active filter in wheel B | |
GPS-INT = ' ' / GPS trigger interval (msec) | Set by the user on the GPS tab of the web interface. This is the
cadence at which GPS pulses are sent. Each pulse triggers a readout on the camera. |
GPSSTART= ' ' / GPS start time (UTC; external) | Set by the user on the GPS tab of the web interface. This is the time
that the first GSP pulse was sent, which corresponds to the readout of the first GPS image. We typically toss the first two images in a GPS triggered sequence because the first is nothing (before the first pulse) and the second has slightly inaccurate timing. |
HA = 'NA ' / The telescope hour angle | Read from the TCS. |
INSTANGL= 'NA ' / The instrument angle | Read from the TCS. |
HA = 'NA ' / The telescope hour angle | Read from the TCS. |
INSTANGL= 'NA ' / The instrument angle | Read from the TCS. |
INSTRUME= 'SHD ' / The SHOC instrument name | |
INSTSWV = '1.0 ' / The SHOC software version | |
OBJDEC = ' ' / Declination of object | From the user-entered information on the advance tab of the web interface. |
OBJECT = ' ' / IAU name of observed object | From the user-entered information on the advance tab of the web interface. |
OBJEPOCH= ' ' / Object coordinate epoch | From the user-entered information on the advance tab of the web interface. |
OBJEQUIN= ' ' / Object coordinate equinox | From the user-entered information on the advance tab of the web interface. |
OBJRA = ' ' / Right ascension of object | From the user-entered information on the advance tab of the web interface. |
OBSERVER= ' ' / Observer who acquired the data | From the user-entered information on the advance tab of the web interface. |
OBSTYPE = ' ' / Observation type | From the user-entered information on the advance tab of the web interface. |
TELDEC = 'NA ' / The telescope declination | Read from the TCS. Currently only on the 74in! |
TELESCOP= '40in ' / The telescope name | |
TELFOCUS= 'NA ' / The telescope focus | Read from the TCS. Currently only on the 74in! |
TELRA = 'NA ' / The telescope right ascension | Read from the TCS. Currently only on the 74in! |
ZD = 'NA ' / The telescope zenith distance | Read from the TCS. Currently only on the 74in! |
END |
FITS Header Keyword Conversion
With the release of the new, web-based software there have been changes in the naming of some of the FITS keywords. The keywords are documented in the table below:
Old Keyword | New Keyword |
---|---|
HIERARCH EMREALGAIN | EMREALGN |
HIERARCH COUNTCONVERTMODE | CNTCVTMD |
HIERARCH COUNTCONVERT | CNTCVT |
HIERARCH DETECTIONWAVELENGTH | DTNWLGTH |
HIERARCH SENSITIVITY | SNTVTY |
HIERARCH SPURIOUSNOISEFILTER | SPSNFLTR |
HIERARCH THRESHOLD | THRSHLD |
HIERARCH PHOTONCOUNTINGENABLED | PCNTENLD |
HIERARCH NOTHRESHOLDS | NSETHSLD |
HIERARCH PHOTONCOUNTINGTHRESHOLD1 | PTNTHLD1 |
HIERARCH PHOTONCOUNTINGTHRESHOLD2 | PTNTHLD2 |
HIERARCH PHOTONCOUNTINGTHRESHOLD3 | PTNTHLD3 |
HIERARCH PHOTONCOUNTINGTHRESHOLD4 | PTNTHLD4 |
HIERARCH AVERAGINGFILTERMODE | AVGFTRMD |
HIERARCH AVERAGINGFACTOR | AVGFCTR |
HIERARCH FRAMECOUNT | FRMCNT |
A script is provided to convert back and forth between the two versions of the keywords. This is useful when working with a data pipeline that is still based on the old version of the keywords or to make old data compatible with a new data pipeline. The script can be used as follows from a command line terminal:
$ convert_keywords.py mydatacube.fits
This will convert the keywords in an old data cube to the new version. To convert the keywords back to the old version, run the script with the "revert" option:
$ convert_keywords.py --revert mydatacube.fits
In both cases the program will try to save the updated cube to a file with the same name, but postfixed with "_converted". So, "mydatacube_converted.fits" in the examples above. If the file already exists, you will be asked if you want to overwrite the file. If you do not want to overwrite the file you will be given the opportunity to enter a different file name. Multiple files can be converted at once by providing a list of them when running the program:
$ convert_keywords.py cube1.fits cube2.fits cube3.fits
NOTE: The conversion script makes use of Astropy. See the official documentation for installation instructions.
In-depth Information
Characteristics and Properties
The specification sheets for the cameras, other manuals and documentation, can be found at http://shoc.saao.ac.za/Documentation.html .
Choosing an Observing Method
- BINNING: The plate scale is 0.076 and 0.163 "/pixel for the 1.9m without and with the focal reducer. It is 0.167"/pixel for the 1.0m without the focal reducer. The optimal binning is ~ 3 pixels per full width half maximum. The seeing conditions in Sutherland are typically > 1"; therefore, SHOC should usually be binned at least 2x2.
- For example, if the seeing is 2" and SHOC is mounted on the 1.9m telescope without the focal reducer, sources would be well sampled at a binning of 6x6.
- SUBFRAMING: The SHOC field of view is not large (1.3' and 2.8' on each side on the 1.9m without and with a focal reducer; 2.9' on each side on the 1.0m with a focal reducer). Subframing is typically only used if the cadence needs to be increased.
- MODE: For most observations, the mode should be selected that allows the desired cadence with the lowest read noise. Start by setting 1 MHz Conventional mode in the optimal binning, and type 0 in the exposure time -- the time will default to the minimum allowable (and the cadence will be displayed). If the cadence is too slow, either subframe or select 3 MHz Conventional mode. The EM mode should only be used in cases of experienced observers who understand both the risks and the noise vs. signal payoffs.
- GAIN: The three gain settings correspond to different electron/ADU conversion values and effect dynamic range. They are slightly different from camera to camera. The standard options for conventional modes are as follows:
- Mode...................... Approx. Gain setting................... Approx. Electrons/ADU
- 1 MHz Conv....................1, 2.5, 5....................................... 4, 1.5, 0.7
- 3 MHz Conv...................1, 2.5, 5....................................... 10, 4, 1.8
User's troubleshooting guide
- Camera won't switch on: If the camera display times out when trying to switch on, you will need to restart the services. To do this:
- Open a terminal. This is available from the top menu or side menu of the thin client in the warm room. If the web browser is full screen, push 'F11', and you should then see the top menu bar. Once the terminal window appears, ssh to the instrument by running the command
ssh shoc74in@shoc74in.suth.saao.ac.za
or
ssh shoc40in@shoc40in.suth.saao.ac.za
When prompted for the password, use the password for the 74in or 40in user.
- In the terminal, type
restart_services
You will again be prompted for the password. The various services comprising the instrument will then restart. This lasts a few seconds a few seconds.
- Return to the web browser, and reload the page, then try to start the camera again.
- It may be necessary to restart the services a few times before you can start the camera.
- Acquisition doesn't start: Check on the GPS tab that the trigger signal has been activated. If it has, but the camera is still not taking images, try stopping the camera, resetting the GPS trigger to a time in the near future, and try again. Sometimes, it appears that the GPS trigger signal is not being sent (despite the software reporting that it is). The current acquisition should be stopped and restarted (because when the acquisition is started, the time the POP signal will start is recorded and will apear in the fits headers - so stopping and starting the acquisition will ensure that the correct POP time is used). Note that the index willl be incremented when doing this. To avoid this, you can rewind the increment by 1, in the advanced tab. Make sure you have no data that will be overwritten.
- Acquisition doesn't finish: Sometimes it appears that the camera does not ever leave the acquiring state (despite all images in the kinetic series having been obtained). It may in this case be neccessary to restart the server, manually edit the FITS files to include the filter and timing informations, and to then copy them to their final destination. The server may be restarted running the command "restart_services" from a terminal. The data is spooled to a directory under /data/spool/, based on the date, the sequence number and the time the acquisition started. In this directory is the data file, data.fits, and file containing other acquisition parameters, info.txt. The latter may be used to update the FITS header. The procedure is as follows:
- Rescue your data using the data rescue utility, as described in the section Rescuing Data.
- Run the command to restart the webserver, as described above in the subsection "Camera won't switch on".
- Return to the web browser, refresh the page, turn the camera back on, and continue with your observing. If the camera is non-responsive, run the restart_services command again.
- IMPORTANT:
- Check that the file number is correct --- look at the Advanced Tab, and the Start Index should be that for the next file. Currently, for the Phase 1 software, this number seems to be the same as the file you just saved from spool and you will likely need to increase it by one to avoid overwriting.
- The filter wheel needs to be reinitialized. Go to the filter wheel tab, click the initialize button under the active wheel(s).esh the camera tab in the browser window
- Camera tab is frozen: Occasionally (often at the start of the night) the first acquisition causes the camera tab to become unresponsive and 'freeze up'. The solution to this is to restart the services, as described above, then refresh the browser window and try again to restart the camera. This may need to be repeated several times before the camera starts.
Technician's guide
Mounting SHOC
- The SHOC box should be mounted at the base of the telescope, and the camera should be mounted below the filter wheel box.
- With the power OFF, connect the following cables:
- computer to camera operation (thick, white cable)
- computer to camera trigger (thin back with gold SMB)
- camera power (black)
- GPS antenna (BNC from telescope into the GPS in the SHOC box)
- network cable (ethernet from back of SHOC box)
- SHOC box power (standard three-prong cable from back of the box)
- the filter wheel to the computer
The power must be off when disconnecting and reconnecting cables.
- To power up:
- (1) Turn on power to the outlet where the SHOC box cable is plugged
- (2) Turn on power switch at the back of the SHOC box.
- (3) Press the computer power button on the front of the computer.
Testing the components
- Once the machine has booted, fire up a browser and point it at http://shocnawe.suth.saao.ac.za:5000/ or http://shocndisbelief.suth.saao.ac.za:5000/.
- In the filterwheel tab:
- Check that the filterwheel(s) can be initialized and moved by selecting the initialize button, then selecting a different filter(s) and moving the wheel(s).
- In the GPS tab:
- Check that the time is valid (it will read "valid" in green). (Note: the time will NOT be valid if the GPS antenna cannot see the sky. The dome must be open for the GPS to see satellites.)
- Check that POP settings can be made:
- Set mode to repeat.
- Pulse width should 10 microseconds (default)
- Start date should be set or selected to be today's date.
- Start time should be set to some seconds or a min in the future, localtime.
- Repeat interval can be your choice: this is just a test value, something like a few hundred millisec is appropriate.
- Click the apply button .
- The status should change to pending, and at the chosen time, should change to activated. Make sure to switch off the POP signal again by clicking stop.
- Make sure to close the dome before trying the camera tests, or else the images could be saturated.
- In the camera tab:
- Check that the camera can be switched on (press "turn camera on")
- Check that the cooler is working (the temperature should drop)
- Check that the camera is reading out: confirm the setting is full frame, conventional mode (the default) and press "preview". Images should appear every ~ 1.2 sec. Once confirmed that images are coming through, press "stop".
- Switch the camera off again (press "turn camera off").
Configuring the SHOC computer (not the web interface!)
To access the SHOC computer, ssh into it from a terminal on your local machine or tablet (e.g. ssh ccd@shocnawe.suth.saao.ac.za or ssh ccd@shocndisbelief.suth.saao.ac.za).
- Ensure GPS location settings are correct in /home/ccd/programming/shoc/shoc/instruments/gps/gps.py:
- after sshing, move into the directory: cd /home/ccd/programming/shoc/shoc/instruments/gps/
- open the file using a viewer: vi gps.py
- Use the up and down arrows to view each line of the file. Check that the location settings for Sutherland are active (not commented with a '#') and that the settings for Cape Town are commented (i.e. are preceded by a '#'). To remove a character type x . To add a character type a followed by the character to add, then hit esc to stop inserting. To exit, type :wq .
- Edit service settings using the file /etc/init/shoc-camera.conf (the filepaths need to have the correct telescope and instrument names):
- after sshing, move into the directory: cd /etc/init/
- open the file using a viewer: vi shoc-camera.conf
- Use the up and down arrows to view each line of the file. Change the telescope name to the name of the currently used telescope: 74in or 40in. The instrument name should already be set to SHA or SHD (for shocnawe or shocndisbelief) and the log level should be info (not debug). In the editor, to remove a character type x . To add a character type a followed by the character to add, then hit esc to stop inserting. To exit, type :wq .
- To enact any changes made during configuration, restart all services:
- Under /home/ccd/programming/shoc, type the following: ./restart_services
Technician's Troubleshooting guide
- SHOC computers have two disks, a solid state drive to which data are spooled and a hard drive which runs the system. If there are concerns about the disks, you can verify mount points:
- At the command line, type the command 'mount'. The solid state disk should be mounted on /data/spool, while the spinning hard disk should be mounted on /data.
- Time is not valid:
- Check that the antenna cable has been connected to the SHOC GPS on the box.
- Check that the antenna has access to the sky (i.e. the dome is open)
- Check that the location settings are correct. Click on 'receiver status' in the GPS tab and look at the latitude and longitude. The Sutherland Latitude is ~32 22 S while Cape Town is 33 56S. If the location is not corrrect, then the GPS will not find satellites and the correct location needs to be updated as described in the Configuring SHOC section.
- Camera is in an inoperable state:
- Restart the services by sshing into the SHOC computer (see Configuring SHOC section):
- change into directory /home/ccd/programming/shoc and type the following: ./restart_services. It may be neccessary to do this a few times.
- If that does not work, under /home/ccd/programming/shoc type the following to stop and start in separate steps:
- Restart the services by sshing into the SHOC computer (see Configuring SHOC section):
- ./stop_services
- ./start_services
- If prompted for the password, it is the same as that for the ccd user.
- Worst case scenario, when restarting the entire computer seems like an appropriate option:
- ssh into the SHOC computer (see Configuring SHOC section)
- type the following: sudo shutdown -h now
- The computer will shutdown. To power back on, someone must MANUALLY go press the button on the front of the SHOC computer.
* If, for whatever reason, you decide to open the SHOC computer case, you MUST USE ELECTROSTATIC PROTECTION. The computer components, especially the PCI card, are electrostatically sensitive. The computer should be placed on a grounding mat and a grounding wriststrap should be used before anything is touched.