=='''2.2  Setting prism position & determining wavelength range '''==
+
=='''Setting prism position & determining wavelength range'''==
  
Do this at the start of the run, using options available by clicking on the “Init” button at the top right of the QUARTZ window:
+
First determine the required prism position for your desired wavelength coverage. 
  
1) Select “Init” → “Find reference” and wait for “Finding reference...” window to close
+
Then '''at the start of the run or after a software crash''', follow these steps using options available by clicking on the “Init” button at the top right of the QUARTZ window:
 +
 
 +
1) Initialise the prisms: select “Init” → “Find reference” and wait for “Finding reference...” window to close
  
 
2) Select “Init” → “Define red” or “Define blue” depending on the prism you wish to use and enter prism position
 
2) Select “Init” → “Define red” or “Define blue” depending on the prism you wish to use and enter prism position
Line 71: Line 73:
 
4) Check and note down the prism position in the “Prism” status box at the bottom of the QUARTZ window.
 
4) Check and note down the prism position in the “Prism” status box at the bottom of the QUARTZ window.
  
5) Take a [[fibreflat|fibre flat]]
+
5) Take a fibre flat and an arc exposure (follow steps in Calibrations section).
 +
 
 +
6) On ltsp, run the data reduction software on the flat & arc files to determine the orders present in the image and the wavelength range.
 +
 
  
6) By taking an arc (follow steps 1-2 in Section 3.3), check the resulting orders are correct (see section on reductions software) and spectrograph focus is good (see Section 5).
+
'''If the calculated orders are far from those expected''', the prisms will need initialising.  Repeat steps (1)-(6) until the expected orders are in the frame.  When this happens, the correct prism position is usually achieved over 3 initialisations, but it is important to move the prism after an initialisation, i.e. always follow steps (1) with step (3) (you can skip step (2) because you already defined your desired position); don't do step (1) twice in a row, or GIRAFFE will get quite upset.
  
 
=='''Calibrations'''==
 
=='''Calibrations'''==

Revision as of 01:15, 28 January 2015

Brief description of GIRAFFE

The SAAO Grating Instrument for Radiation Analysis with a Fibre Fed Échelle - GIRAFFE - consists of two components:

(i) The head which is mounted at the Cassegrain focus to collect light from the star and direct it into the fibre;

(ii) The spectrograph, constructed on an optical bench in the Coude room, in which the light emerging from the fibre is dispersed and recorded by a CCD camera.

The CCD camera is controlled by a PC in the observing room. The software, called QUARTZ, has the task of setting the exposure time, reading and displaying the CCD image and also controlling various components on the GIRAFFE head, such as the arc lamp and flatfield lamps. At the end of the exposure the FITS file is transferred to the server and can be displayed and reduced via the thin client.

There are four PCs that the observer will use:

1. A Linux PC called giraffe running QUARTZ.

2. A thin client system for data storage ltsp.suth.saao.ac.za.

3. A laptop running a LabView program to display the APD counts, i.e. monitor the light intensity emerging from the fibre.

4. The Linux TCS PC for acquisition and autoguiding.

Data reduction Software for Giraffe

Filling the cryostat

After filling the cryostat, you must wait 30 mins before beginning observations.

A technician will fill the cryostat at ~08:00 and ~16:00 (check on the log outside the Coude room that this has been done before you start observing). The observer must fill the cryostat either at ~19:00 and at the end of the night (avoiding the need to stop during observing), or at ~00:00, whether or not observations have been made.

The cryostat is filled from the dewar outside the Coude room. Check the pressure of the dewar - it should be ~10 bar, and must not exceed 15 bar. Open the right-hand blue tap to start filling, and watch the cryostat. Close the tap when liquid nitrogen starts to drip out of the cryostat.

The CCD "tank" temperature should be ~180K, and the "sink" temperature ~79K. The temperatures are displayed toward the bottom of the QUARTZ window.

Even in bad weather, the cryostat must be filled during the night. The last filling for the night should not be done before midnight.

Instrument & software startup

The instrument is controlled by the QUARTZ software on giraffe PC.

2.1 Starting the software

1) Log on as ccd (usual password)

2) cd to working directory

3) Enter command: new_quartz

4) In the pop-up CCD menu, select MUS1

5) In the second pop-up window, select Local setup files → select giraffe.girReadyOk


Each GIRAFFE run is assigned a run number, that QUARTZ will now prompt you to enter. To determine your run number, look in the GIRAFFE log book located in the 1.9-m dome book rack and look for the last entry. Increase that run number by one. If you are a returning observer and wish to use the same directory, you will need to change your run number by deleting the file disk.file from your directory.


QUARTZ will now initialise the CCD controller and check the status of the GIRAFFE head. Do the following:

  • Wait 2 minutes for it to finish initalising.
  • Check that things are working, by reading out the CCD by selecting Read CCD from the CCD drop-down menu at the top right of the QUARTZ window. This will display information regarding the bias reading and the maximum CCD count.
  • Select “Filter” → “Init” at start of run, and again before every filter change

Setting prism position & determining wavelength range

First determine the required prism position for your desired wavelength coverage.

Then at the start of the run or after a software crash, follow these steps using options available by clicking on the “Init” button at the top right of the QUARTZ window:

1) Initialise the prisms: select “Init” → “Find reference” and wait for “Finding reference...” window to close

2) Select “Init” → “Define red” or “Define blue” depending on the prism you wish to use and enter prism position

3) Select “Init” → “Red/blue position go” and wait for “prism movement” and “detector motion” messages to clear

4) Check and note down the prism position in the “Prism” status box at the bottom of the QUARTZ window.

5) Take a fibre flat and an arc exposure (follow steps in Calibrations section).

6) On ltsp, run the data reduction software on the flat & arc files to determine the orders present in the image and the wavelength range.


If the calculated orders are far from those expected, the prisms will need initialising. Repeat steps (1)-(6) until the expected orders are in the frame. When this happens, the correct prism position is usually achieved over 3 initialisations, but it is important to move the prism after an initialisation, i.e. always follow steps (1) with step (3) (you can skip step (2) because you already defined your desired position); don't do step (1) twice in a row, or GIRAFFE will get quite upset.

Calibrations

3.1 Camera Flats

Taking camera flats involves interacting with the components on the bench in the Coude room. Please be extremely careful not to touch anything except the flatfield screen and the lamp.

At the start of the run, or after a software crash, obtain camera flats as follows:

1) Enter the Coudé room, duck under the liquid nitrogen pipe and lift the large black cover over the side of the bench

2) Holding the handle of the white screen, slide it in front of the camera.

3) Go to the opposite side of the bench, switch on the small lamp from the shelf below and turn the intensity to very low. Carefully place the lamp directly in front of the screen, so as to illuminate it as uniformly as possible.

4) In the QUARTZ “Control” panel, select “ND Pos 1” and “Mir Pos 3”.

5) In the bottom left window of the QUARTZ software, type *CAMERA

6) Click the “Expose” button in the “Control” panel on the right of the QUARTZ window and enter the exposure time (likely a couple of seconds for red prism). Counts should be as high as possible (<50000) without saturating.

7) After readout, check the information box on the bottom left – if saturated pixels are reported, reduce the exposure time. Flats with saturated pixels cannot be used. If satisfied, save to disk.

8) Check image is transferred to the thin client ltsp.suth:/data/74in/giraffe/data/image/yyyymmdd, and check header info to ensure it is your file (not an old file with the same name). If data is not found on ltsp, check /data/image/yyyymmdd on the giraffe PC, and call IT (**113) to set up image transfer to ltsp.

9) Expose and save 10 camera flats.

10) Carefully remove the lamp and screen.


Fibre Flats

At the start of an observation, take ~7 fibre flats as follows:

1) Rotate the silver knob on the GIRAFFE head on the telescope to the IN position, to insert a colour filter into the beam.

2) Select “ND Pos 1”, “Mir Pos 3” and “Lamp” in the QUARTZ “Control” panel.

3) Click the “Expose” button in the “Control” panel on the right of the QUARTZ window and enter the exposure time. Counts should be as high as possible (<50000) without saturating. Depending on prism setting, it may be necessary to use a different ND filter.

4) After readout, check the counts in the image and the information box on the bottom left – if saturated pixels are reported, reduce the exposure time.

5) If satisfied, save to disk. Expose and save 3 fibre flats.

6) On QUARTZ, click to switch OFF “Lamp”. At the telescope, rotate the silver knob on the GIRAFFE head to the OUT position.


3.3 Arcs

At the start of an observation, take 2 arc exposures as follows:

1) Select “ND Pos 1”, “Mir Pos 3” and “Arc” in the QUARTZ “Control” panel

2) Click the “Expose” button in the “Control” panel on the right of the QUARTZ window and enter the exposure time. It is ok to saturate a few pixels of an arc exposure.

3) Switch on APD: click “APD” button in QUARTZ “Control” panel.

4) After readout, check counts on the image & in the QUARTZ info box on the bottom left. Counts should be as high as possible (<50000).

5) If satisfied, save to disk. Expose & save 2 arcs. Switch OFF “Arc”.

6) Switch OFF the APD when slewing/taking camera or fibre flats.

Observing

4.1 Target exposures

1) Select “ND Pos 1” and “Mir Pos 1” in the QUARTZ “Control” panel

2) Type “*target_name” in the text box at the bottom left (e.g. *TPyx)

3) Click the “Expose” button in the “Control” panel on the right of the QUARTZ window and enter the exposure time.

4) Take an arc every ~20 minutes, or on acquiring a new target.